Use of a Mixing Chamber for Sample Preparation and Multiple Collection in Sequential Injection Analysis: Determination of Sulfate in Wines

View document

Bibliographic Details
Summary:A sequential injection system for determination of sulfate in wines was developed. It was based on the formation of BaSO4, measured turbidimetrically at 420 nm. As sample dilution and acidification were required, both sample and acid were sent to a mixing chamber placed in a lateral port of the selection valve; after thorough mixing, up to six aliquots could be drawn and analyzed. A linear calibration curve was established using 10% v/v ethanol standards with concentrations between 300 and 1500 mg L-1 of K2SO4. The results obtained by the present methodology were statistically comparable with those furnished by the reference procedure, with relative deviations < 3.5%. The sample frequency was about 5 samples per h with relative standard deviations (RSD) < 10%.
Main Authors:Silva, Helena R.
Other Authors:Segundo, Marcela A.; Rangel, António O. S. S.
Subject:Sequential injection Mixing chamber Turbidimetry sulfate Wine
Year:2003
Country:Portugal
Document type:article
Access type:open access
Associated institution:Universidade Católica Portuguesa
Language:English
Origin:Veritati - Repositório Institucional da Universidade Católica Portuguesa
Description
Summary:A sequential injection system for determination of sulfate in wines was developed. It was based on the formation of BaSO4, measured turbidimetrically at 420 nm. As sample dilution and acidification were required, both sample and acid were sent to a mixing chamber placed in a lateral port of the selection valve; after thorough mixing, up to six aliquots could be drawn and analyzed. A linear calibration curve was established using 10% v/v ethanol standards with concentrations between 300 and 1500 mg L-1 of K2SO4. The results obtained by the present methodology were statistically comparable with those furnished by the reference procedure, with relative deviations < 3.5%. The sample frequency was about 5 samples per h with relative standard deviations (RSD) < 10%.

Funded activities

No funded projects found.