Publicação
The influence of different screw tightening forces on the vertical misfit of implant-supported frameworks
| Resumo: | OBJECTIVES: The present in vitro study was designed to compare the differences in the vertical misfit of implant-supported frameworks using three different forces for tightening the bridge locking screws: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. METHODS: The investigation was conducted based on the results given by 9 six-unit nickel-chromium (2 abutments/ 4 pontics) screw-retained implant-supported frameworks. The structures were exposed to simulated porcelain firings. The marginal misfit measurements were made using a traveling measuring microscope at selected screw tightening forces: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. The results were submitted to one-way ANOVA with repeated measures on one factor, and post hoc pairwise comparisons using Tukey test (5%). RESULTS: The mean marginal misfit of the frameworks, fastening the screws by hand until first resistance, was 41.56µm (SD±12.45µm). The use of torque driver devices caused a significant reduction in marginal opening (p<0.05). With the lowest torque available (10Ncm), the mean marginal discrepancy at the abutment-framework interface was reduced an average of 52% to a mean marginal opening of 19.71µm (SD±2.97µm). After the use of the 20Ncm torque driver, the mean marginal discrepancy of the frameworks was reduced an average of 69% to a mean marginal opening of 12.82µm (SD±4.0µm). Comparing the use of torque drivers with 10 and 20 Ncm torque, the means are not significantly different from one another. CONCLUSION: The seating force has an important effect on the vertical misfit measurements, once it may considerably narrow the vertical misfit gaps at the abutment-framework interface, thus leading to a misjudgment of the real marginal situation. |
|---|---|
| Autores principais: | Vasconcellos,Diego Klee de |
| Outros Autores: | Bottino,Marco Antonio; Nishioka,Renato Sussumu; Valandro,Luiz Felipe; Costa,Elza Maria Valadares da |
| Assunto: | Dental implants Osseointegration Implant-supported dental prosthesis Torque misfit |
| Ano: | 2005 |
| País: | Brasil |
| Tipo de documento: | artigo |
| Tipo de acesso: | acesso aberto |
| Instituição associada: | IBICT |
| Idioma: | inglês |
| Origem: | Oasisbr |
| Resumo: | OBJECTIVES: The present in vitro study was designed to compare the differences in the vertical misfit of implant-supported frameworks using three different forces for tightening the bridge locking screws: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. METHODS: The investigation was conducted based on the results given by 9 six-unit nickel-chromium (2 abutments/ 4 pontics) screw-retained implant-supported frameworks. The structures were exposed to simulated porcelain firings. The marginal misfit measurements were made using a traveling measuring microscope at selected screw tightening forces: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. The results were submitted to one-way ANOVA with repeated measures on one factor, and post hoc pairwise comparisons using Tukey test (5%). RESULTS: The mean marginal misfit of the frameworks, fastening the screws by hand until first resistance, was 41.56µm (SD±12.45µm). The use of torque driver devices caused a significant reduction in marginal opening (p<0.05). With the lowest torque available (10Ncm), the mean marginal discrepancy at the abutment-framework interface was reduced an average of 52% to a mean marginal opening of 19.71µm (SD±2.97µm). After the use of the 20Ncm torque driver, the mean marginal discrepancy of the frameworks was reduced an average of 69% to a mean marginal opening of 12.82µm (SD±4.0µm). Comparing the use of torque drivers with 10 and 20 Ncm torque, the means are not significantly different from one another. CONCLUSION: The seating force has an important effect on the vertical misfit measurements, once it may considerably narrow the vertical misfit gaps at the abutment-framework interface, thus leading to a misjudgment of the real marginal situation. |
|---|
Atividades financiadas
Não foram encontrados projetos financiados.