Multichannel photomultiplier for multipass thomson scattering diagnostics
| Resumo: | The application of a new multichannel 64 parallel amplification 10 stage photomultiplier arranged in 8 x 8 matrix of 20.32 mm x 20.32 mm, detector XP1752 from Philips, for laser light scattering diagnostics is now under development. Signal intensity calibration curve for eight parallel and individual channels, as well as cross talk between channels was measured. Also, Rayleigh scattering on nitrogen gas, normally used as optical calibration method for electron density measurements in Thomson scattering was realized, and the intensity of both detectors was compared making measurements with a tungsten lamp. |
|---|---|
| Autores principais: | Monteiro,M. J. R. |
| Outros Autores: | Machida,M.; Daltrini,A. M.; Berni,L. A. |
| Ano: | 2002 |
| País: | Brasil |
| Tipo de documento: | artigo |
| Tipo de acesso: | acesso aberto |
| Instituição associada: | IBICT |
| Idioma: | inglês |
| Origem: | Oasisbr |
| Resumo: | The application of a new multichannel 64 parallel amplification 10 stage photomultiplier arranged in 8 x 8 matrix of 20.32 mm x 20.32 mm, detector XP1752 from Philips, for laser light scattering diagnostics is now under development. Signal intensity calibration curve for eight parallel and individual channels, as well as cross talk between channels was measured. Also, Rayleigh scattering on nitrogen gas, normally used as optical calibration method for electron density measurements in Thomson scattering was realized, and the intensity of both detectors was compared making measurements with a tungsten lamp. |
|---|