Development of sequential injection methodologies for the spectrophotometric direct and kinetic determination of aluminium in natural and waste waters

Ver documento

Detalhes bibliográficos
Resumo:Two sequential injection methodologies for the spectrophotometric determination of reactive aluminium in water samples are proposed, based on the reaction between chrome azurol S (CAS) and aluminium. The two methods involve a direct and a kinetic methodology (for its application to coloured waste waters), both accommodated within the same manifold. The working ranges obtained were: 0.040-0.500 mg Al L-1 for direct method and 0.050-0.300 mg Al L-1 for the kinetic method. The described system has high degree of automation, enabling sampling throughputs of 31 and 57 h-1 for the kinetic method and direct methods, respectively. The reagents consumption per determination is low: 46 µg of CAS, 3.75 mg of sodium acetate and 25 µg of ascorbic acid. The volume of effluent produced per determination is only around 2.8 mL in the direct method and 3.4 mL in the kinetic method.
Autores principais:Mesquita,Raquel B. R.
Outros Autores:Rangel,António O. S. S.
Assunto:sequential injection spectrophotometry aluminium waters kinetic determination
Ano:2008
País:Brasil
Tipo de documento:artigo
Tipo de acesso:acesso aberto
Instituição associada:IBICT
Idioma:inglês
Origem:Oasisbr
Descrição
Resumo:Two sequential injection methodologies for the spectrophotometric determination of reactive aluminium in water samples are proposed, based on the reaction between chrome azurol S (CAS) and aluminium. The two methods involve a direct and a kinetic methodology (for its application to coloured waste waters), both accommodated within the same manifold. The working ranges obtained were: 0.040-0.500 mg Al L-1 for direct method and 0.050-0.300 mg Al L-1 for the kinetic method. The described system has high degree of automation, enabling sampling throughputs of 31 and 57 h-1 for the kinetic method and direct methods, respectively. The reagents consumption per determination is low: 46 µg of CAS, 3.75 mg of sodium acetate and 25 µg of ascorbic acid. The volume of effluent produced per determination is only around 2.8 mL in the direct method and 3.4 mL in the kinetic method.

Atividades financiadas

Não foram encontrados projetos financiados.