Resumo:The performance of axial view ICP OES and ICP-MS techniques were compared through the determination of As, Fe, Mn, Pb, Sb and Sn in electrolytic copper. Samples were prepared by two procedures: 1. Total dissolution with 5 % v/v HNO3 and 2. Dissolution with 1.4 % v/v HNO3 plus 2.0 % v/v H2SO4, followed by separation of the Cu by electrodeposition. The methods were applied for the analysis of standard copper solutions, using calibration against aqueous solutions with or without the addition of Cu. The results obtained by the two techniques were similar, but the separation of Cu from the sample matrix proved to be more adequate for practical reasons.
Autores principais:Santos,Éder José dos
Outros Autores:Herrmann,Amanda Beatriz; Olkuszewski,José Luiz; Saint'Pierre,Tatiana D.; Curtius,Adilson José
Assunto:ICP-MS axial view ICP OES electrolytic copper electrodeposition
Ano:2005
País:Brasil
Tipo de documento:artigo
Tipo de acesso:acesso aberto
Instituição associada:IBICT
Idioma:inglês
Origem:Oasisbr
Descrição
Resumo:The performance of axial view ICP OES and ICP-MS techniques were compared through the determination of As, Fe, Mn, Pb, Sb and Sn in electrolytic copper. Samples were prepared by two procedures: 1. Total dissolution with 5 % v/v HNO3 and 2. Dissolution with 1.4 % v/v HNO3 plus 2.0 % v/v H2SO4, followed by separation of the Cu by electrodeposition. The methods were applied for the analysis of standard copper solutions, using calibration against aqueous solutions with or without the addition of Cu. The results obtained by the two techniques were similar, but the separation of Cu from the sample matrix proved to be more adequate for practical reasons.

Atividades financiadas