Smart textile for respiratory monitoring and thoraco-abdominal motion pattern evaluation

View document

Bibliographic Details
Summary:The use of wearable systems for monitoring vital parameters has gained wide popularity in several medical fields. The focus of the present study is the experimental assessment of a smart textile based on 12 fiber Bragg grating sensors for breathing monitoring and thoraco‐abdominal motion pattern analysis. The feasibility of the smart textile for monitoring several temporal respiratory parameters (ie, breath‐by‐breath respiratory period, breathing frequency, duration of inspiratory and expiratory phases), volume variations of the whole chest wall and of its compartments is performed on 8 healthy male volunteers. Values gathered by the textile are compared to the data obtained by a motion analysis system, used as the reference instrument. Good agreement between the 2 systems on both respiratory period (bias of 0.01 seconds), breathing frequency (bias of −0.02 breaths/min) and tidal volume (bias of 0.09 L) values is demonstrated. Smart textile shows good performance in the monitoring of thoraco‐abdominal pattern and its variation, as well.
Main Authors:Silvatti, Amanda P.
Other Authors:Massaroni, Carlo; Venanzi, Cecilia; Presti, Daniela Lo; Saccomandi, Paola; Formica, Domenico; Giurazza, Francesco; Caponero, Michele A.; Schena, Emiliano
Subject:Breathing monitoring Fiber Bragg gratings Fiber optics Smart textiles Wearables
Year:2018
Country:Brazil
Document type:article
Access type:open access
Associated institution:IBICT
Language:English
Origin:Oasisbr
Description
Summary:The use of wearable systems for monitoring vital parameters has gained wide popularity in several medical fields. The focus of the present study is the experimental assessment of a smart textile based on 12 fiber Bragg grating sensors for breathing monitoring and thoraco‐abdominal motion pattern analysis. The feasibility of the smart textile for monitoring several temporal respiratory parameters (ie, breath‐by‐breath respiratory period, breathing frequency, duration of inspiratory and expiratory phases), volume variations of the whole chest wall and of its compartments is performed on 8 healthy male volunteers. Values gathered by the textile are compared to the data obtained by a motion analysis system, used as the reference instrument. Good agreement between the 2 systems on both respiratory period (bias of 0.01 seconds), breathing frequency (bias of −0.02 breaths/min) and tidal volume (bias of 0.09 L) values is demonstrated. Smart textile shows good performance in the monitoring of thoraco‐abdominal pattern and its variation, as well.

Funded activities

No funded projects found.